#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2304T (Expiry date: 24.09.2024)
Production date : 25.01.2023
Results of quality control 2 (after
6 months), date: 24.07.2023
No serum blank (average) : 0.125
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.008 |
-0.010 |
0.016 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.012 |
0.021 |
0.003 |
0.046 |
0.021 |
1 |
Weak positive serum
(N° 9200-07) |
1.384 |
1.447 |
1.482 |
1.501 |
1.454 |
63 |
Positive control serum
(N° 9200-08) |
2.230 |
2.274 |
2.382 |
2.339 |
2.306 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.306 |
Yes |
Absorbance of weak positive serum 9200-07, as % of positive serum |
> 32 % |
63 % |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 6 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.125 |
Yes |
Comments:
Good for use.