Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2303C (Expiry date: 04.09.2024)

Production date : 05.01.2023

Results of quality control 3 (after 12 months), date: 04.01.2024

No serum blank (average) : 0.178

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.007 0.007 -0.012 0.012 0.000 0
Negative control serum
(N° 9700-06)
0.089 0.155 0.068 0.121 0.108 5
Weak positive serum
(N° 9700-07)
0.386 0.445 0.412 0.440 0.421 19
Positive control serum
(N° 9700-08)
2.078 2.439 2.126 2.330 2.243 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.243 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 19 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 5 % Yes
Absorbance of no serum blank < 0.350 0.178 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland