#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2303C (Expiry date: 04.09.2024)

Production date : 05.01.2023

Results of quality control 1 (after production), date: 11.01.2023

No serum blank (average) : 0.178

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.014 0.004 -0.017 -0.001 0.000 0
Negative control serum
(N° 9700-06)
0.110 0.119 0.048 0.079 0.089 4
Weak positive serum
(N° 9700-07)
0.444 0.483 0.384 0.351 0.415 18
Positive control serum
(N° 9700-08)
2.244 2.442 2.249 2.332 2.316 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.316 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 18 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.178 Yes

Comments:

Good for use.