Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2302C (Expiry date: 04.09.2024)

Production date : 05.01.2023

Results of quality control 1 (after production), date: 11.01.2023

No serum blank (average) : 0.191

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 -0.006 0.011 -0.005 0.000 0
Negative control serum
(N° 9700-06)
0.117 0.103 0.109 0.104 0.109 5
Weak positive serum
(N° 9700-07)
0.449 0.448 0.455 0.454 0.452 19
Positive control serum
(N° 9700-08)
2.325 2.400 2.255 2.461 2.361 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.361 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 19 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 5 % Yes
Absorbance of no serum blank < 0.350 0.191 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch