#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of echinococcoses in humans
Lot 2238G
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus granulosus IgG ELISA, Article 9350
Expiry date: 21.07.2024, Control date: 16.11.2023,
Control # 3
No serum blank (average) : 0.115
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
-0.002 |
0.004 |
-0.002 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.067 |
0.069 |
0.062 |
0.080 |
0.070 |
3 |
Weak positive serum
(N° 9350-07) |
1.193 |
1.196 |
1.219 |
1.229 |
1.209 |
53 |
Positive control serum
(N° 9350-08) |
2.146 |
2.344 |
2.176 |
2.462 |
2.282 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.282 |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 12 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.115 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +