Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2236H

 

Results of quality control

Kit: Bordier Affinity Products SA, Entamoeba histolytica IgG ELISA, Article 9550.
Expiry date: 21.07.2024, Control date: 19.05.2023,  Control #  2

No serum blank (average) : 0.124

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.000 -0.004 -0.001 0.005 0.000 0
Negative control serum
(N° 9550-06)
0.166 0.162 0.154 0.155 0.159 6
Weak positive serum
(N° 9550-07)
0.608 0.637 0.747 0.739 0.683 26
Positive control serum
(N° 9550-08)
2.463 2.465 2.729 2.818 2.619 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.619 Yes
Absorbance of negative serum 9550-06, as % of positive serum < 14 % 6 % Yes
Absorbance of no serum blank < 0.350 0.124 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch