Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 2232M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis (Em2-Em18) IgG ELISA, Article 9300
Expiry date: 16.06.2024, Control date: 25.10.2022,  Control #  1

No serum blank (average) : 0.132

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 -0.005 0.001 0.007 0.000 0
Negative control serum
(N° 9300-06)
0.032 0.017 0.041 0.034 0.031 1
Weak positive serum
(N° 9300-07)
0.650 0.651 0.681 0.685 0.666 29
Positive control serum
(N° 9300-08)
2.250 2.372 2.261 2.298 2.295 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.295 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 12 % 1 % Yes
Absorbance of no serum blank < 0.350 0.132 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch