#6100
Aspergillus
fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis
Lot 2231A
Results of quality control
Kit: Bordier Affinity Products SA, Aspergillus fumigatus IgG ELISA, Article
6100 .
Expiry date: 16.06.2024, Control date: 18.04.2023,
Control # 2
No serum blank (average) : 0.111
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.001 |
-0.004 |
0.002 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 6100-06) |
0.061 |
0.065 |
0.049 |
0.066 |
0.060 |
3 |
Weak positive serum
(N° 6100-07) |
0.420 |
0.439 |
0.414 |
0.432 |
0.426 |
20 |
Positive control
serum
(N° 6100-08) |
2.035 |
2.210 |
2.082 |
2.052 |
2.095 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 6100-08 |
> 1.200 |
2.095 |
Yes |
Absorbance of negative serum 6100-06, as % of positive serum |
< 10 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.111 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|