#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2230D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica IgG ELISA, Article
9650 .
Expiry date: 04.06.2024, Control date: 05.06.2024,
Control # 4
No serum blank (average) : 0.113
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
-0.003 |
0.003 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.115 |
0.100 |
0.088 |
0.103 |
0.102 |
4 |
Weak positive serum
(N° 9650-07) |
0.411 |
0.418 |
0.421 |
0.383 |
0.409 |
17 |
Positive control
serum
(N° 9650-08) |
2.314 |
2.481 |
2.343 |
2.419 |
2.390 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.390 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.113 |
Yes |
Comments:
Good for use.