#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2230D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica IgG ELISA, Article
9650 .
Expiry date: 04.06.2024, Control date: 29.09.2023,
Control # 3
No serum blank (average) : 0.119
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.002 |
-0.004 |
0.001 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.096 |
0.125 |
0.108 |
0.115 |
0.111 |
5 |
Weak positive serum
(N° 9650-07) |
0.420 |
0.440 |
0.413 |
0.434 |
0.427 |
17 |
Positive control
serum
(N° 9650-08) |
2.382 |
2.570 |
2.385 |
2.524 |
2.466 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.466 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
5 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.119 |
Yes |
Comments:
Good for use.