#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2230D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica IgG ELISA, Article
9650 .
Expiry date: 04.06.2024, Control date: 11.10.2022,
Control # 1
No serum blank (average) : 0.118
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.003 |
-0.002 |
0.000 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.104 |
0.092 |
0.116 |
0.103 |
0.104 |
4 |
Weak positive serum
(N° 9650-07) |
0.445 |
0.447 |
0.480 |
0.500 |
0.468 |
19 |
Positive control
serum
(N° 9650-08) |
2.321 |
2.511 |
2.456 |
2.430 |
2.430 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.430 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.118 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|