#9310
Echinococcus
multilocularis (Em18) IgG ELISA
ELISA kit for alveolar echinococcosis patient follow-up in humans
Lot 2228F
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus multilocularis (Em18)
IgG ELISA, Article 9310
Expiry date: 04.06.2024, Control date: 12.10.2022,
Control # 1
No serum blank (average) : 0.118
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
-0.003 |
0.005 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9310-06) |
0.008 |
0.011 |
0.029 |
0.020 |
0.017 |
1 |
Weak positive serum
(N° 9310-07) |
0.144 |
0.161 |
0.168 |
0.175 |
0.162 |
7 |
Positive control serum
(N° 9310-08) |
2.381 |
2.408 |
2.482 |
2.479 |
2.437 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9310-08 |
> 1.200 |
2.437 |
Yes |
Absorbance of negative serum 9310-06, as % of positive serum |
< 5 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.118 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|