#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2225T
Results of quality control
Kit: Bordier Affinity Products SA, Toxocara canis IgG ELISA, Article 9200
Expiry date: 14.05.2024, Control date: 13.09.2023,
Control # 3
No serum blank (average) : 0.116
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.001 |
-0.002 |
0.002 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.037 |
0.026 |
0.015 |
0.015 |
0.023 |
1 |
Weak positive serum
(N° 9200-07) |
1.581 |
1.517 |
1.539 |
1.598 |
1.559 |
64 |
Positive control serum
(N° 9200-08) |
2.408 |
2.623 |
2.227 |
2.558 |
2.454 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.454 |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 8 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.113 |
Yes |
Comments:
Good for use.