#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2225T
Results of quality control
Kit: Bordier Affinity Products SA, Toxocara canis IgG ELISA, Article 9200
Expiry date: 14.05.2024, Control date: 14.03.2023,
Control # 2
No serum blank (average) : 0.113
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
-0.004 |
0.003 |
0.003 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.025 |
0.013 |
0.011 |
0.023 |
0.018 |
1 |
Weak positive serum
(N° 9200-07) |
1.498 |
1.520 |
1.509 |
1.494 |
1.505 |
60 |
Positive control serum
(N° 9200-08) |
2.470 |
2.592 |
2.506 |
2.469 |
2.509 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.509 |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 8 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.113 |
Yes |
Comments:
Good for use.