#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of trichinellosis in humans
Lot 2220R
Results of quality control
Kit: Bordier Affinity Products SA, Trichinella spiralis IgG ELISA, Article
9750 .
Expiry date: 11.03.2024, Control date: 05.07.2023,
Control # 3
No serum blank (average) : 0.119
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
-0.003 |
0.006 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.047 |
0.047 |
0.049 |
0.051 |
0.048 |
2 |
Weak positive serum
(N° 9750-07) |
0.462 |
0.490 |
0.461 |
0.457 |
0.467 |
20 |
Positive control serum
(N° 9750-08) |
2.254 |
2.393 |
2.318 |
2.478 |
2.360 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9750-08 |
> 1.200 |
2.360 |
Yes |
Absorbance of negative serum 9750-06, as % of positive serum |
< 10 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.119 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|