Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9250 Ascaris IgG ELISA
ELISA kit for the diagnosis of ascariasis in humans

Lot 2214U

 

Results of quality control

Kit: Bordier Affinity Products SA, Ascaris IgG ELISA, Article 9250.
Expiry date: 18.01.2024, Control date: 27.05.2022,  Control #  1

No serum blank (average) : 0.115

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.004 -0.005 0.003 -0.002 0.000 0
Negative control serum
(N° 9250-06)
0.086 0.080 0.074 0.077 0.079 4
Weak positive serum
(N° 9250-07)
0.684 0.647 0.668 0.664 0.665 30
Positive control serum
(N° 9250-08)
2.127 2.236 2.191 2.255 2.202 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9250-08 > 1.200 2.202 Yes
Absorbance of negative serum 9250-06, as % of positive serum < 10 % 4 % Yes
Absorbance of no serum blank < 0.350 0.115 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch