Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2212H

 

Results of quality control

Kit: Bordier Affinity Products SA, Entamoeba histolytica IgG ELISA, Article 9550.
Expiry date: 10.12.2023, Control date: 10.10.2022,  Control #  2

No serum blank (average) : 0.142

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.001 -0.007 0.004 0.001 0.000 0
Negative control serum
(N° 9550-06)
0.086 0.105 0.139 0.124 0.114 5
Weak positive serum
(N° 9550-07)
0.457 0.462 0.470 0.481 0.468 22
Positive control serum
(N° 9550-08)
2.078 2.133 2.180 2.255 2.162 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.162 Yes
Absorbance of negative serum 9550-06, as % of positive serum < 14 % 5 % Yes
Absorbance of no serum blank < 0.350 0.142 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland