#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2211D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica IgG ELISA, Article
9650 .
Expiry date: 10.12.2023, Control date: 10.10.2022,
Control # 2
No serum blank (average) : 0.116
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.001 |
0.002 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.078 |
0.094 |
0.087 |
0.087 |
0.086 |
4 |
Weak positive serum
(N° 9650-07) |
0.365 |
0.390 |
0.389 |
0.378 |
0.380 |
17 |
Positive control
serum
(N° 9650-08) |
2.164 |
2.337 |
2.231 |
2.271 |
2.250 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.250 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.116 |
Yes |
Comments:
Good for use.