#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2209T
Results of quality control
Kit: Bordier Affinity Products SA, Toxocara canis IgG ELISA, Article 9200
Expiry date: 20.11.2023, Control date: 15.03.2023,
Control # 3
No serum blank (average) : 0.126
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.002 |
-0.007 |
0.008 |
-0.004 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.025 |
0.009 |
0.021 |
0.023 |
0.020 |
1 |
Weak positive serum
(N° 9200-07) |
1.234 |
1.180 |
1.314 |
1.313 |
1.261 |
59 |
Positive control serum
(N° 9200-08) |
2.047 |
2.096 |
2.135 |
2.328 |
2.152 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.152 |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 8 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.126 |
Yes |
Comments:
Good for use.