Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2108H

 

Results of quality control

Kit: Bordier Affinity Products SA, Entamoeba histolytica IgG ELISA, Article 9550.
Expiry date: 10.11.2022, Control date: 19.03.2021,  Control #  1

No serum blank (average) : 0.153

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.001 -0.006 0.010 -0.006 0.000 0
Negative control serum
(N° 9550-06)
0.096 0.119 0.108 0.125 0.112 5
Weak positive serum
(N° 9550-07)
0.495 0.525 0.518 0.518 0.514 24
Positive control serum
(N° 9550-08)
2.136 2.175 2.158 2.192 2.166 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.166 Yes
Absorbance of negative serum 9550-06, as % of positive serum < 14 % 5 % Yes
Absorbance of no serum blank < 0.350 0.153 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch