Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#6100 Aspergillus fumigatus
ELISA kit for the diagnosis of human aspergillosis

Lot 2006A

 

Results of quality control

Kit: Bordier Affinity Products SA, Aspergillus fumigatus ELISA, Article 6100.
Expiry date: 18.10.2021, Control date: 16.02.2021,  Control #  3

No serum blank (average) : 0.139

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 0.000 0.003 -0.002 0.000 0
Negative control serum
(N° 61
00-06)
0.032 0.068 0.079 0.090 0.067 3
Weak positive serum
(N° 61
00-07)
0.644 0.624 0.678 0.613 0.640 33
Positive control serum
(N° 61
00-08)
1.951 1.933 1.963 1.943 1.948 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 1.948 Yes
Absorbance of negative serum 6100-06, as % of positive serum < 10 % 3 % Yes
Absorbance of no serum blank < 0.350 0.139 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch