#6100
Aspergillus
fumigatus
ELISA kit for the diagnosis of human aspergillosis
Lot 2006A
Results of quality control
Kit: Bordier Affinity Products SA, Aspergillus fumigatus ELISA, Article
6100 .
Expiry date: 18.10.2021, Control date: 17.08.2020,
Control # 2
No serum blank (average) : 0.131
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.000 |
0.002 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 6100-06) |
0.050 |
0.062 |
0.047 |
0.062 |
0.055 |
3 |
Weak positive serum
(N° 6100-07) |
0.646 |
0.690 |
0.615 |
0.641 |
0.648 |
32 |
Positive control
serum
(N° 6100-08) |
1.967 |
2.120 |
1.906 |
2.026 |
2.005 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 6100-08 |
> 1.200 |
2.005 |
Yes |
Absorbance of negative serum 6100-06, as % of positive serum |
< 10 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.131 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|