#9350
Echinococcus
granulosus
ELISA kit for the diagnosis of echinococcoses in humans
Lot 1927G
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus granulosus ELISA, Article 9350
Expiry date: 02.08.2021, Control date: 27.11.2020,
Control # 3
No serum blank (average) : 0.135
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
0.002 |
0.004 |
-0.004 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.050 |
0.053 |
0.041 |
0.057 |
0.050 |
2 |
Weak positive serum
(N° 9350-07) |
1.001 |
1.053 |
1.051 |
1.101 |
1.047 |
49 |
Positive control serum
(N° 9350-08) |
1.954 |
2.268 |
2.021 |
2.302 |
2.136 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.136 |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 12 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.135 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|