#9400 Acanthocheilonema viteae (Onchocercidae)
ELISA kit for the diagnosis of human filariasis

Lot 1925V

Results of quality control

Kit: Bordier Affinity Products SA, Acanthocheilonema viteae ELISA, Article 9400.
Expiry date: 28.07.2021, Control date: 06.12.2019,  Control #  1

No serum blank (average) : 0.136

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.007 -0.001 0.001 0.007 0.000 0
Negative control serum
(N° 9400-06)
0.051 0.088 0.091 0.091 0.080 3
Weak positive serum
(N° 9400-07)
0.847 0.890 0.945 0.881 0.891 38
Positive control serum
(N° 9400-08)
2.342 2.344 2.357 2.305 2.337 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.337 Yes
Absorbance of negative serum 9400-06, as % of positive serum < 12 % 3 % Yes
Absorbance of no serum blank < 0.350 0.136 Yes

Comments:

Good for use.