#9650
Fasciola
hepatica (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 1921D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica ELISA, Article
9650 .
Expiry date: 11.05.2021, Control date: 07.09.2020,
Control # 3
No serum blank (average) : 0.127
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
0.003 |
0.002 |
-0.002 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.059 |
0.087 |
0.083 |
0.076 |
0.076 |
4 |
Weak positive serum
(N° 9650-07) |
0.284 |
0.271 |
0.278 |
0.269 |
0.275 |
14 |
Positive control
serum
(N° 9650-08) |
1.968 |
2.062 |
1.968 |
2.097 |
2.023 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.023 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.127 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +