#9300
Echinococcus
multilocularis (Em2-Em18)
ELISA kit for the diagnosis of alveolar echinococcosis in humans
Lot 1920M
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus multilocularis
Em2-Em18 ELISA, Article
9300
Expiry date: 11.05.2021, Control date: 07.09.2020,
Control # 3
No serum blank (average) : 0.132
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
-0.002 |
0.001 |
0.003 |
0.000 |
0 |
Negative control serum
(N° 9300-06) |
0.013 |
0.019 |
0.029 |
0.033 |
0.024 |
1 |
Weak positive serum
(N° 9300-07) |
0.624 |
0.611 |
0.646 |
0.613 |
0.624 |
29 |
Positive control serum
(N° 9300-08) |
2.065 |
2.142 |
2.139 |
2.171 |
2.129 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9300-08 |
> 1.200 |
2.129 |
Yes |
Absorbance of negative serum 9300-06, as % of positive serum |
< 12 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.132 |
Yes |
Comments:
Good for use.