Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#6100 Aspergillus fumigatus
ELISA kit for the diagnosis of human aspergillosis

Lot 1907A

 

Results of quality control

Kit: Bordier Affinity Products SA, Aspergillus fumigatus ELISA, Article 6100.
Expiry date: 05.11.2020, Control date: 02.03.2020,  Control #  3

No serum blank (average) : 0.130

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.003 0.001 -0.001 -0.003 0.000 0
Negative control serum
(N° 61
00-06)
0.043 0.031 0.044 0.042 0.040 2
Weak positive serum
(N° 61
00-07)
0.490 0.463 0.467 0.437 0.464 28
Positive control serum
(N° 61
00-08)
1.690 1.714 1.619 1.611 1.659 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 1.659 Yes
Absorbance of negative serum 6100-06, as % of positive serum < 10 % 2 % Yes
Absorbance of no serum blank < 0.350 0.130 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch