Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9600 Schistosoma mansoni
ELISA kit for the diagnosis of human schistosomiasis

Lot 1902B

 

Results of quality control

Kit: Bordier Affinity Products SA, Schistosoma mansoni ELISA, Article 9600.
Expiry date: 02.09.2020, Control date: 03.09.2020,  Control #  4

No serum blank (average) : 0.131

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.000 -0.005 0.002 0.002 0.000 0
Negative control serum
(N° 9600-06)
0.025 0.048 0.025 0.049 0.037 2
Weak positive serum
(N° 9600-07)
0.330 0.312 0.352 0.357 0.338 16
Positive control serum
(N° 9600-08)
1.907 2.090 2.005 2.229 2.058 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.058 Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.131 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland