#9350
Echinococcus
granulosus
ELISA kit for the diagnosis of echinococcoses in humans
Lot 1901G
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus granulosus ELISA, Article 9350
Expiry date: 02.09.2020, Control date: 03.09.2020,
Control # 4
No serum blank (average) : 0.132
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.007 |
-0.003 |
0.006 |
0.004 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.028 |
0.053 |
0.039 |
0.058 |
0.044 |
2 |
Weak positive serum
(N° 9350-07) |
0.916 |
0.900 |
0.937 |
1.050 |
0.950 |
44 |
Positive control serum
(N° 9350-08) |
1.937 |
2.172 |
2.095 |
2.362 |
2.141 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.141 |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 12 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.132 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +