Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9650 Fasciola hepatica (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 1828D

 

Results of quality control

Kit: Bordier Affinity Products SA, Fasciola hepatica ELISA, Article 9650.
Expiry date: 22.07.2020, Control date: 05.12.2018,  Control #  1

No serum blank (average) : 0.123

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.005 -0.005 0.006 0.004 0.000 0
Negative control serum
(N° 965
0-06)
0.108 0.148 0.172 0.130 0.139 6
Weak positive serum
(N° 965
0-07)
0.399 0.392 0.430 0.390 0.402 17
Positive control serum
(N° 9650
-08)
2.244 2.385 2.406 2.544 2.394 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.394 Yes
Absorbance of negative serum 9650-06, as % of positive serum < 9 % 6 % Yes
Absorbance of no serum blank < 0.350 0.123 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch