Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#6100 Aspergillus fumigatus
ELISA kit for the diagnosis of human aspergillosis

Lot 1804A

 

Results of quality control

Kit: Bordier Affinity Products SA, Aspergillus fumigatus ELISA, Article 6100.
Expiry date: 08.09.2019, Control date: 09.07.2018,  Control #  2

No serum blank (average) : 0.145

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 -0.002 0.003 0.002 0.000 0
Negative control serum
(N° 61
00-06)
0.063 0.064 0.047 0.056 0.058 3
Weak positive serum
(N° 61
00-07)
0.668 0.635 0.715 0.706 0.681 32
Positive control serum
(N° 61
00-08)
1.988 2.080 2.180 2.213 2.116 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.116 Yes
Absorbance of negative serum 6100-06, as % of positive serum < 10 % 3 % Yes
Absorbance of no serum blank < 0.350 0.145 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch