#9600 Schistosoma mansoni
ELISA kit for the diagnosis of human schistosomiasis

Lot 1802B

 

Results of quality control

Kit: Bordier Affinity Products SA, Schistosoma mansoni ELISA, Article 9600.
Expiry date: 08.09.2019, Control date: 07.01.2019,  Control #  3

No serum blank (average) : 0.182

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.001 -0.005 0.003 0.002 0.000 0
Negative control serum
(N° 9600-06)
0.051 0.050 0.038 0.048 0.046 2
Weak positive serum
(N° 9600-07)
0.336 0.385 0.328 0.329 0.344 17
Positive control serum
(N° 9600-08)
2.012 2.149 1.982 2.161 2.076 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.076 Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.182 Yes

Comments:

Good for use.