Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9600 Schistosoma mansoni
ELISA kit for the diagnosis of human schistosomiasis

Lot 1802B

 

Results of quality control

Kit: Bordier Affinity Products SA, Schistosoma mansoni ELISA, Article 9600.
Expiry date: 08.09.2019, Control date: 06.07.2018,  Control #  2

No serum blank (average) : 0.167

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.012 -0.003 0.011 0.004 0.000 0
Negative control serum
(N° 9600-06)
0.032 0.034 0.048 0.058 0.043 2
Weak positive serum
(N° 9600-07)
0.398 0.409 0.426 0.423 0.414 20
Positive control serum
(N° 9600-08)
1.966 2.052 2.006 2.236 2.065 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.065 Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.167 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch