#9400 Acanthocheilonema viteae (Onchocercidae)
ELISA kit for the diagnosis of human filariasis

Lot 1725V

Results of quality control

Kit: Bordier Affinity Products SA, Acanthocheilonema viteae ELISA, Article 9400.
Expiry date: 19.05.2019, Control date: 13.09.2018,  Control #  3

No serum blank (average) : 0.160

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 -0.003 0.001 0.004 0.000 0
Negative control serum
(N° 9400-06)
0.087 0.098 0.112 0.120 0.104 4
Weak positive serum
(N° 9400-07)
0.840 0.830 0.872 0.944 0.871 34
Positive control serum
(N° 9400-08)
2.450 2.527 2.487 2.816 2.570 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.570 Yes
Absorbance of negative serum 9400-06, as % of positive serum < 12 % 4 % Yes
Absorbance of no serum blank < 0.350 0.160 Yes

Comments:

Good for use.