Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9300 Echinococcus multilocularis (Em2-Em18)
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1721M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis Em2-Em18 ELISA, Article 9300
Expiry date: 16.04.2019, Control date: 17.04.2019,  Control #  4

No serum blank (average) : 0.175

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.013 -0.012 0.003 -0.004 0.000 0
Negative control serum
(N° 9300-06)
0.073 0.099 0.066 0.103 0.085 3
Weak positive serum
(N° 9300-07)
0.742 0.766 0.794 0.794 0.774 31
Positive control serum
(N° 9300-08)
2.332 2.495 2.444 2.609 2.470 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.470 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 3 % Yes
Absorbance of no serum blank < 0.350 0.175 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland