#9300 Echinococcus multilocularis (Em2-Em18)
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1721M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis Em2-Em18 ELISA, Article 9300
Expiry date: 16.04.2019, Control date: 14.08.2018,  Control #  3

No serum blank (average) : 0.144

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.001 -0.005 0.007 -0.003 0.000 0
Negative control serum
(N° 9300-06)
0.069 0.071 0.088 0.063 0.072 3
Weak positive serum
(N° 9300-07)
0.631 0.657 0.640 0.614 0.635 30
Positive control serum
(N° 9300-08)
2.042 2.203 1.981 2.118 2.086 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.086 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 3 % Yes
Absorbance of no serum blank < 0.350 0.144 Yes

Comments:

Good for use.