#9300 Echinococcus multilocularis (Em2-Em18)
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1721M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis Em2-Em18 ELISA, Article 9300
Expiry date: 16.04.2019, Control date: 12.02.2018,  Control #  2

No serum blank (average) : 0.160

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 -0.001 0.001 0.001 0.000 0
Negative control serum
(N° 9300-06)
0.078 0.092 0.091 0.104 0.091 4
Weak positive serum
(N° 9300-07)
0.682 0.634 0.651 0.683 0.663 30
Positive control serum
(N° 9300-08)
2.164 2.232 2.159 2.370 2.231 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.231 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 4 % Yes
Absorbance of no serum blank < 0.350 0.160 Yes

Comments:

Good for use.