Echinococcus multilocularis
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1721M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis ELISA, Article 9300
Expiry date: 16.04.2019, Control date: 23.08.2017,  Control #  1

No serum blank (average) : 0.171

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.007 0.004 -0.007 -0.003 0.000 0
Negative control serum
(N° 9300-06)
0.086 0.104 0.087 0.085 0.090 4
Weak positive serum
(N° 9300-07)
0.857 0.827 0.789 0.793 0.816 33
Positive control serum
(N° 9300-08)
2.509 2.468 2.459 2.363 2.450 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.450 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 4 % Yes
Absorbance of no serum blank < 0.350 0.171 Yes

Comments:

Good for use.