Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Aspergillus fumigatus
ELISA kit for the diagnosis of human aspergillosis

Lot 1709A

 

Results of quality control

Kit: Bordier Affinity Products SA, Aspergillus fumigatus ELISA, Article 6100.
Expiry date: 05.12.2018, Control date: 26.04.2016,  Control #  1

No serum blank (average) : 0.141

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.006 0.002 -0.001 0.005 0.000 0
Negative control serum
(N° 61
00-06)
0.076 0.075 0.088 0.092 0.082 3
Weak positive serum
(N° 61
00-07)
0.986 0.940 0.981 0.956 0.965 40
Positive control serum
(N° 61
00-08)
2.461 2.363 2.473 2.358 2.413 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.413 Yes
Absorbance of negative serum 6100-06, as % of positive serum < 10 % 3 % Yes
Absorbance of no serum blank < 0.350 0.141 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland