Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Fasciola hepatica (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 1708D

 

Results of quality control

Kit: Bordier Affinity Products SA, Fasciola hepatica ELISA, Article 9650.
Expiry date: 31.10.2018, Control date: 03.03.2017,  Control #  1

No serum blank (average) : 0.145

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.003 -0.004 0.005 0.002 0.000 0
Negative control serum
(N° 965
0-06)
0.115 0.129 0.140 0.136 0.130 5
Weak positive serum
(N° 965
0-07)
0.391 0.396 0.410 0.433 0.408 17
Positive control serum
(N° 9650
-08)
2.266 2.504 2.324 2.541 2.409 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.409 Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 5 % Yes
Absorbance of no serum blank < 0.350 0.145 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch