Echinococcus multilocularis
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1706M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis ELISA, Article 9300
Expiry date: 31.10.2018, Control date: 28.08.2017,  Control #  2

No serum blank (average) : 0.172

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.002 -0.007 0.003 0.003 0.000 0
Negative control serum
(N° 9300-06)
0.087 0.110 0.101 0.117 0.103 4
Weak positive serum
(N° 9300-07)
0.781 0.774 0.784 0.820 0.789 32
Positive control serum
(N° 9300-08)
2.402 2.468 2.365 2.587 2.455 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.455 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 4 % Yes
Absorbance of no serum blank < 0.350 0.172 Yes

Comments:

Good for use.