Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Entamoeba histolytica
ELISA kit for the diagnosis of human visceral amebiasis

Lot 1704H

 

Results of quality control

Kit: Bordier Affinity Products SA, Entamoeba histolytica ELISA, Article 9550.
Expiry date: 09.09.2018, Control date: 13.01.2017,  Control #  1

No serum blank (average) : 0.157

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.009 -0.006 0.003 0.012 0.000 0
Negative control serum
(N° 9550-06)
0.214 0.201 0.207 0.214 0.209 9
Weak positive serum
(N° 9550-07)
0.661 0.620 0.687 0.746 0.678 29
Positive control serum
(N° 9550-08)
2.254 2.119 2.361 2.610 2.336 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.336 Yes
Absorbance of negative serum 9550-06, as % of positive serum < 14 % 9 % Yes
Absorbance of no serum blank < 0.350 0.157 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch