Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Fasciola hepatica (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 1628D

 

Results of quality control

Kit: Bordier Affinity Products SA, Fasciola hepatica ELISA, Article 9650.
Expiry date: 06.08.2018, Control date: 07.08.2018,  Control #  4

No serum blank (average) : 0.145

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.003 0.002 -0.003 0.004 0.000 0
Negative control serum
(N° 965
0-06)
0.143 0.168 0.145 0.167 0.155 7
Weak positive serum
(N° 965
0-07)
0.324 0.353 0.311 0.363 0.337 15
Positive control serum
(N° 9650
-08)
2.152 2.408 2.125 2.345 2.257 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.257 Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 7 % Yes
Absorbance of no serum blank < 0.350 0.145 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch