Kit: Bordier Affinity Products SA, Schistosoma mansoni ELISA, Article 9600.
Expiry date: 23.06.2018, Control date: 17.10.2017,  Control #  3

No serum blank (average) : 0.156

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.001 -0.008 0.004 0.004 0.000 0
Negative control serum
(N° 9600-06)
0.098 0.084 0.076 0.087 0.086 4
Weak positive serum
(N° 9600-07)
0.417 0.409 0.438 0.423 0.421 21
Positive control serum
(N° 9600-08)
1.985 2.005 2.039 2.146 2.043 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.043 Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.156 Yes

Comments:

Good for use.