Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Schistosoma mansoni
ELISA kit for the diagnosis of human schistosomiasis

Lot 1625B

 

Results of quality control

Kit: Bordier Affinity Products SA, Schistosoma mansoni ELISA, Article 9600.
Expiry date: 23.06.2018, Control date: 21.04.2017,  Control #  2

No serum blank (average) : 0.153

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.003 -0.004 0.005 0.002 0.000 0
Negative control serum
(N° 9600-06)
0.098 0.091 0.093 0.108 0.097 4
Weak positive serum
(N° 9600-07)
0.460 0.459 0.460 0.466 0.461 21
Positive control serum
(N° 9600-08)
2.159 2.245 2.226 2.318 2.237 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.237 Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.153 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch