Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis ELISA, Article 9300
Expiry date: 27.05.2018, Control date: 28.05.2018,  Control #  4

No serum blank (average) : 0.197

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.008 0.008 -0.001 0.002 0.000 0
Negative control serum
(N° 9300-06)
0.091 0.094 0.071 0.088 0.086 4
Weak positive serum
(N° 9300-07)
0.673 0.642 0.697 0.703 0.679 28
Positive control serum
(N° 9300-08)
2.286 2.365 2.284 2.619 2.388 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.388 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 4 % Yes
Absorbance of no serum blank < 0.350 0.197 Yes

Comments:

Good for use.