Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Echinococcus multilocularis
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1623M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis ELISA, Article 9300
Expiry date: 27.05.2018, Control date: 22.09.2017,  Control #  3

No serum blank (average) : 0.163

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.002 0.001 0.000 -0.002 0.000 0
Negative control serum
(N° 9300-06)
0.109 0.108 0.080 0.098 0.099 4
Weak positive serum
(N° 9300-07)
0.745 0.790 0.591 0.721 0.712 27
Positive control serum
(N° 9300-08)
2.569 2.789 2.490 2.638 2.621 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.621 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 4 % Yes
Absorbance of no serum blank < 0.350 0.163 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch