Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

Strongyloides ratti
ELISA kit for the diagnosis of human strongyloidosis

Lot 1617S

Results of quality control

Kit: Bordier Affinity Products SA, Strongyloides ratti ELISA. Article 9450.
Expiry date: 10.03.2018, Control date: 07.07.2017,  Control #  3

No serum blank (average) : 0.155

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.000 0.000 0.001 -0.002 0.000 0
Negative control serum
(N° 9450-06)
0.089 0.100 0.112 0.108 0.103 5
Weak positive serum
(N° 9450-07)
0.695 0.708 0.759 0.747 0.728 33
Positive control serum
(N° 9450-08)
2.058 2.182 2.187 2.309 2.184 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.184 Yes
Absorbance of negative serum 9450-06, as % of positive serum < 12 % 5 % Yes
Absorbance of no serum blank < 0.350 0.155 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch