Fasciola hepatica (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 1615D

 

Results of quality control

Kit: Bordier Affinity Products SA, Fasciola hepatica ELISA, Article 9650.
Expiry date: 06.02.2018, Control date: 02.06.2017,  Control #  3

No serum blank (average) : 0.139

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
-0.004 0.001 0.004 -0.001 0.000 0
Negative control serum
(N° 965
0-06)
0.172 0.174 0.170 0.160 0.169 8
Weak positive serum
(N° 965
0-07)
0.381 0.323 0.314 0.317 0.333 15
Positive control serum
(N° 9650
-08)
2.133 2.319 2.217 2.299 2.242 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.242 Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 8 % Yes
Absorbance of no serum blank < 0.350 0.139 Yes

Comments:

The negative serum 9650-06 just exceeds its specification of <8%. Further experiments have been made to insure that it was a ponctual non coformity. The threshold and so the diagnostic value of the product are not affected, nethertheless the product is under watching for future quality controls.
Good for use.