Echinococcus multilocularis
ELISA kit for the diagnosis of alveolar echinococcosis in humans

Lot 1606M

Results of quality control

Kit: Bordier Affinity Products SA, Echinococcus multilocularis ELISA, Article 9300
Expiry date: 21.10.2017, Control date: 27.10.2017,  Control #  4

No serum blank (average) : 0.167

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
(TBS-Tween)
0.005 -0.003 -0.001 -0.001 0.000 0
Negative control serum
(N° 9300-06)
0.129 0.132 0.130 0.134 0.131 5
Weak positive serum
(N° 9300-07)
0.819 0.853 0.808 0.833 0.828 32
Positive control serum
(N° 9300-08)
2.423 2.524 2.508 2.741 2.549 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.549 Yes
Absorbance of negative serum 9300-06, as % of positive serum < 15 % 5 % Yes
Absorbance of no serum blank < 0.350 0.167 Yes

Comments:

Good for use.